Guidelines for Accurate Target Cell Counts Using 10x Genomics® Single Cell Solutions

Guidelines for Accurate Target Cell Counts Using 10x Genomics® Single Cell Solutions

  • CG000091_10x_Technical Note_Guidelines on Accurate Target Cell Count_RevB.pdf

    Technical Note, CG000091

    CG000091_10x_Technical Note_Guidelines on Accurate Target Cell Count_RevB.pdf

    View and download

Technical Note - Guidelines for Accurate Target Cell Counts Using 10x Genomics® Single Cell Solutions

This Technical Note presents an overview of best practices for cell handling and counting that will increase the accuracy of cell counts and minimize the variability in the number of loaded cells. 10x Genomics Single Cell Protocols require suspensions of viable, single cells as input. Accurate counting and qualification of cells during the workflow of the Single Cell 3’ v2 and Single Cell V(D)J Protocol is essential to determine the number of cells to load on to the microfluidic chip for analysis on the system. While manual (e.g. hemocytometer) and automated (e.g. Countess II Automated Cell Counter) counting methods can be used to report cell concentrations, additional factors should be considered that can affect the final count.

FOR USE WITH

  • Chromium™ Single Cell 3’ v2, v3, v3.1, and HT v3.1 products
  • Chromium™ Single Cell 5’ v1, v1.1, v2, and HT v2 products
Important
This technical note presents guidelines for counting cells in the range of 700-1200 cells/µl. If using the Single Cell 3' LT v3.1 (low throughput) application, ensure cells are counted as indicated in this protocol and then diluted to the LT specific optimal loading concentration of 100-600 cells/µl using the Cell Dilution Overview in the LT User Guide.
Document Type
Technical Note