Supported thermal cyclers for Visium workflows
This information was previously located in the Thermal Cycler Recommendations for 10x Genomics Single Cell and Spatial Protocols Technical Note CG000571
Supported thermal cyclers for Xenium workflows
Supported thermal cyclers for Chromium Single Cell workflows
Supported thermal cyclers for Xenium workflows
Supported thermal cyclers for Chromium Single Cell workflows
Thermal Cycler | Part Number | Visium HD | Visium CytAssist Gene Expression | Visium CytAssist Gene & Protein Expression | Visium FFPE v1 | Visium FF v1 |
---|---|---|---|---|---|---|
Analytik Jena Biometra TAdvanced 96 S † | 846-x-070-251 | |||||
Analytik Jena Biometra TAdvanced 96 SG † | 846-x-070-241 | |||||
Eppendorf Mastercycler X50s ‡ | 6311000010 | |||||
Eppendorf Mastercycler X50a ‡ | 6313000018 | |||||
BioRad PTC Tempo Deepwell | 12015392 | |||||
ThermoFisher VeritiPro ¶ | A48141 | |||||
VWR XT96 Gradient | 76452-153 | - | - | - | - | - |
MJ Research PTC-200 (discontinued) | 05434-05 | - | - | - | - | - |
BioRad C1000 (discontinued) | 1851197 | |||||
ThermoFisher Veriti (discontinued) | 4375786 | - | ||||
Eppendorf Mastercycler Pro (discontinued) | 950030010 | - |
Table legend
Supported
Not supported
-
Not tested
Footnote
† Analytik Jena Biometra TAdvanced 96 S/SG: 2°C/sec ramp rates for both heating and cooling. x in part number = 2 for 230 V; 4 for 115 V; 5 for 100 V, 50-60 Hz)
‡ Eppendorf Mastercycler X50s/X50a: 3°C/sec ramp rate for heating and 2°C/sec ramp rate for cooling
¶ ThermoFisher VeritiPro requires FW 1.2.0, 96 well tray/retainer (PN 4381850), and “Cover Ramping” enabled